Disinfecting - B.Braun Aesculap Mode D'emploi/Description Technique

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 38
Aesculap
LED light source OP950 with sterile adapter OP941
8.8.2
Mechanical neutral or mild alkaline cleaning and thermal disinfecting
Machine type: single-chamber cleaning/disinfection device without ultrasound
Phase
Step
I
Prerinse
II
Cleaning
III
Intermediate rinse
IV
Thermal disinfecting
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least)
*Recommended: BBraun Helimatic Cleaner neutral
Check visible surfaces for residues after mechanical cleaning/disinfect-
ing.
Repeat the cleaning/disinfecting process if necessary.
14
®
T
t
Water
[°C/°F]
[min]
quality
<25/77
3
D–W
55/131
10
FD-W
>10/50
1
FD-W
90/194
5
FD-W
-
-
-
Chemical
-
Neutral:
Concentrate:
– pH neutral
– <5 % anionic surfactant
0.5 %* working solution
Mildly alkaline:
Concentrate:
– pH = 9.5
– <5 % anionic surfactant
0.5 % solution
-
-
According to the program for cleaning and disinfection device