Manual cleaning/disinfection

Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
infecting solution.
After manual cleaning/disinfection, check visible surfaces visually for residues.
Repeat the cleaning /disinfection process if necessary.
Manual cleaning with immersion disinfection
Phase
Step
T
[°C/°F]
I
Disinfecting
RT (cold)
cleaning
II
Intermediate
RT (cold)
rinse
III
Disinfection
RT (cold)
IV
Final rinse
RT (cold)
V
Drying
RT
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality
at least)
RT:
Room temperature
*Recommended: BBraun Stabimed
Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure.
Phase I
Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are
moistened.
Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed
from the surface.
If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min.
Mobilize non-rigid components, such as set screws, links, etc. during cleaning.
Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a
disposable syringe.
Phase II
Rinse/flush the product thoroughly (all accessible surfaces) under running water.
Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
Drain any remaining water fully.
Phase III
Fully immerse the product in the disinfectant solution.
Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
Rinse lumens at least 5 times at the beginning of the exposure time using an appropriate disposable syringe.
Ensure that all accessible surfaces are moistened.
Phase IV
Rinse/flush the product thoroughly (all accessible surfaces).
Mobilize non-rigid components, such as set screws, joints, etc. during final rinse.
Rinse lumens with an appropriate disposable syringe at least five times.
Drain any remaining water fully.
Phase V
Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning
and disinfection procedure.

Mechanical cleaning/disinfecting

Note
The cleaning and disinfecting machine must be of tested and approved effectiveness (e.g. FDA approval or CE mark
according to DIN EN ISO 15883).
Note
The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
Mechanical alkaline cleaning and thermal disinfecting
Machine type: single-chamber cleaning/disinfection device without ultrasound
Phase
Step
T
[°C/°F]
I
Prerinse
<25/77
II
Cleaning
55/131
III
Intermediate
>10/50
rinse
IV
Thermal disin-
90/194
fecting
V
Drying
-
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality
at least)
*Recommended: BBraun Helimatic Cleaner alkaline
Check visible surfaces for residues after mechanical cleaning/disinfecting.
t
Conc.
Water
Chemical
[%]
quality
[min]
>15
2
D–W
Aldehyde-free, phenol-free,
and QUAT-free concentrate,
pH ~ 9*
1
-
D–W
-
15
2
D–W
Aldehyde-free, phenol-free,
and QUAT-free concentrate,
pH ~ 9*
1
-
FD-W
-
-
-
-
-
t
Water
Chemical/Note
quality
[min]
3
D–W
-
10
FD-W
Concentrate, alkaline:
– pH = 13
– <5 % anionic surfactant
0.5 % working solution
– pH = 11*
1
FD-W
-
5
FD-W
-
-
-
According to the program for cleaning
and disinfection device
Mechanical cleaning/disinfection with manual pre-cleaning
Note
The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark
according to DIN EN ISO 15883).
Note
The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
Manual pre-cleaning with ultrasound and brush
Phase
Step
T
[°C/°F]
I
Ultrasonic
RT (cold)
cleaning
II
Rinsing
RT (cold)
D–W:
Drinking water
RT:
Room temperature
*Recommended: BBraun Stabimed
Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure.
Phase I
Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible
surfaces are immersed and acoustic shadows are avoided.
Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed
from the surface.
If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min.
Mobilize non-rigid components, such as set screws, links, etc. during cleaning.
Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a
disposable syringe.
Phase II
Rinse/flush the product thoroughly (all accessible surfaces) under running water.
Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
Mechanical alkaline cleaning and thermal disinfecting
Machine type: single-chamber cleaning/disinfection device without ultrasound
Phase
Step
T
[°C/°F]
I
Prerinse
<25/77
II
Cleaning
55/131
III
Intermediate
>10/50
rinse
IV
Thermal disin-
90/194
fecting
V
Drying
-
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality
at least)
*Recommended: BBraun Helimatic Cleaner alkaline
Check visible surfaces for residues after mechanical cleaning/disinfecting.
Inspection, maintenance and checks
Damage (metal seizure/friction corrosion) to the product caused by insufficient
lubrication!
Prior to function checks, lubricate moving parts (e.g. joints, pusher compo-
nents and threaded rods) at the marked lubrication points with maintenance
CAUTION
oil suitable for the respective sterilization process (e.g. for steam sterilization:
Aesculap STERILIT® I oil spray JG600 or STERILIT® I drip lubricator JG598), see
Fig. 2.
Allow the product to cool down to room temperature.
After each complete cleaning, disinfecting and drying cycle, check that the product is dry, clean, operational, and
free of damage (e.g. broken insulation or corroded, loose, bent, broken, cracked, worn, or fractured components).
Dry the product if it is wet or damp.
Repeat cleaning and disinfection of products that still show impurities or contamination.
Check that the product functions correctly.
Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see Tech-
nical Service.
Check for compatibility with associated products.
Packaging
Store products with ratchet locks fully opened or locked no further than in the first notch.
Place the product in its holder or on a suitable tray. Ensure that all cutting edges are protected.
Pack trays appropriately for the intended sterilization process (e.g. in Aesculap sterile containers).
Ensure that the packaging provides sufficient protection against recontamination of the product during storage.

Steam sterilization

Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g. by
opening any valves and faucets).
Validated sterilization process
– Steam sterilization using fractional vacuum process
– Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665
– Sterilization using fractional vacuum process at 134 °C, minimum holding time 5 min
When sterilizing several products at the same time in a steam sterilizer, ensure that the maximum load capacity
of the steam sterilizer specified by the manufacturer is not exceeded.
t
Conc.
Water
Chemical
[%]
quality
[min]
>15
2
D–W
Aldehyde-free, phenol-free,
and QUAT-free concen-
trate, pH ~ 9*
1
-
D–W
-
t
Water
Chemical
quality
[min]
3
D–W
-
10
FD-W
Concentrate, alkaline:
– pH = 13
– <5 % anionic surfactant
0.5 % working solution
– pH = 11*
1
FD-W
-
5
FD-W
-
-
-
According to the program for cleaning
and disinfection device