Automatic Cleaning/Disinfection With Manual Pre-Cleaning; Manual Pre-Cleaning With A Brush - B.Braun Aesculap Acculan 4 GA344 Mode D'emploi/Description Technique

Petite perceuse
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 36
7.7

Automatic cleaning/disinfection with manual pre-cleaning

Note
The cleaning and disinfection device must fundamentally have a tested
efficacy (such as FDA approval or CE label pursuant to DIN EN ISO 15883).
Note
The cleaning and disinfection machine used for processing must be ser-
viced and checked at regular intervals.
7.7.1

Manual pre-cleaning with a brush

Phase
Step
I
Rinsing
II
Brushes
D–W:
Drinking water
RT:
Room temperature
Do not clean the product in a ultrasonic bath and do not immerse the
product in any fluids. Let any fluid incursions drain out immediately,
otherwise there is a danger of corrosion and loss of function.
Phase I
Move flexible components (such as sleeves) during cleaning.
Thoroughly clean the product under running water.
Phase II
Move flexible components (such as sleeves) during cleaning.
Brush cannulation with cleaning brush TA011944 and difficult to
access surfaces with a suitable plastic cleaning brush for at least
1 min.
After manual preparation, check visible surfaces for residue and repeat
the pre-cleaning process as needed.
Note
For details on difficult to access surfaces, see Acculan pre-cleaning and
care information TA016000 (available in Aesculap Extranet at
https://extranet.bbraun.com).
T
t
Conc.
Water
[°C/°F]
[min]
[%]
quality
RT (cold) -
-
D–W
RT (cold) -
-
D–W
Chemistry/Note
Until visually clean
Until visually clean
11

Publicité

Table des Matières
loading

Table des Matières