B.Braun Aesculap Acculan 4 GA344 Mode D'emploi/Description Technique page 17

Petite perceuse
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 36
Issue
Cause
Kirschner wire protection
Kirschner wire protection
sleeve cannot be
sleeve not compatible
attached
Connection to product
deformed/defective
Thread on Kirschner wire pro-
tection sleeve
deformed/defective
Rinsing adapter cannot
Rinsing adapter not compati-
be attached
ble
Connection to product
deformed/defective
Thread on rinsing adapter
deformed/defective
Cap cannot be
Cap not compatible
attached/removed
Cap deformed/defective
Connection to product
deformed/defective
Battery cannot be
Battery not compatible
inserted/removed
Battery deformed/defective
Battery shaft to product
deformed/defective
Sterile funnel or battery
Sterile funnel or battery
removal device not
removal device not compati-
attachable
ble
Sterile funnel or battery
removal device
deformed/defective
Battery shaft to product
deformed/defective
Oil spray adapter not
Oil spray adapter not compat-
attachable
ible
Speed control knob not
Product in locked OFF position Safety catch is in OFF position
working
Speed control knob sticks/is
defective
Left rotation knob not
Left rotation knob sticks/is
working
defective
Detection
Kirschner wire protection sleeve cannot
be screwed on
Kirschner wire protection sleeve cannot
be screwed on
Kirschner wire protection sleeve cannot
be screwed on
Rinsing adapter cannot be screwed on
Rinsing adapter cannot be screwed on
Rinsing adapter cannot be screwed on
Cap does not click into place
Cap is difficult or cannot be
mounted/removed
Cap is difficult or cannot be
mounted/removed
Battery cannot be completely inserted
into the battery shaft
Battery is difficult or cannot be
mounted/removed
Battery is difficult or cannot be
mounted/removed
Sterile funnel or battery removal device
not attachable to the battery shaft
Sterile funnel or battery removal device
difficult or not attachable
Sterile funnel or battery removal device
difficult or not attachable
Oil spray adapter not attachable
Speed control knob not working
Left rotation knob not working
Remedy
Use the compatible Kirschner wire protection
sleeve for GA344.
Have the manufacturer repair the product.
Change Kirschner wire protection sleeve.
Use the compatible rinsing adapter for GA344.
Have the manufacturer repair the product.
Change rinsing adapter.
Use the compatible cap for GA344.
Have the manufacturer repair the cap.
Have the manufacturer repair the product.
Use the compatible battery for GA344.
Have the manufacturer repair the battery.
Have the manufacturer repair the product.
Use compatible sterile funnel or battery
removal device for GA344.
Change sterile funnel or battery device.
Have the manufacturer repair the product.
Use the compatible oil spray adapter for
GA344.
Turn the safety catch into the ON position.
Have the manufacturer repair the product.
Have the manufacturer repair the product.
15

Publicité

Table des Matières
loading

Table des Matières