C) e4000-07: Size L
3.3 mm
2.0 mm
Approximate Span
Before Closure
2.0 mm
The shape of the closed clip may vary according to tissue thickness.
The amount of Titanium that a patient is exposed to is 97 grams per Applier if the entire load
of the applier is implanted.
D) e4000-08: Size XL
4.9 mm
3.0 mm
Approximate Span
Before Closure
3.0 mm
The shape of the closed clip may vary according to tissue thickness.
The amount of Titanium that a patient is exposed to is 140 grams per Applier if the entire load
of the applier is implanted.
WARNING
• SYMMETRICALLY EVERT ALL TISSUE EDGES
• PLACE CLIPS AS CLOSE AS POSSIBLE TO ONE ANOTHER
• AVOID USE ON VESSELS SEVERELY COMPROMISED DURING ENDARTERECTOMY
FAILURE TO OBSERVE THE ABOVE MAY CONTRIBUTE TO INTEROPERATIVE OR POST-OPERATIVE
(SEVERAL DAYS) FAILURE OF ANASTOMOSIS RESULTING IN SERIOUS PATIENT INJURY.
Cautions
1. Squeeze the levers together fully until a discernible click is felt. Failure to squeeze the levers
completely can result in clip malformation and possible bleeding or leakage.
2. Ensure that the tissue to be anastomosed fi ts completely within the confi nes of the jaws or bleeding
and leakage may result. See fi gure below.
3. Place the AnastoClips in such a fashion that they are not "rocking" on their axis ("tips"). See fi gure
below.
4. Inspect the anastomotic site to ensure proper application and that hemostasis has been achieved.
If bleeding is observed after application, additional clips or placement of manual sutures may be
necessary to complete hemostasis.
5. Do not evert the tissue by grasping one tissue edge with one pair of forceps and the other tissue
edge with another pair of forceps and pulling them together to evert and apply the clips. This may
result in asymmetrical eversion of tissue, which could result in possible bleeding or leakage.
6. Inspect the tissue wall to ensure that the forceps do not damage tissue during manipulation.
7. When using the AnastoClip applier with tissue or prosthetic graft material (e.g., Meadox, knitted
polyester) ensure that the total thickness of the everted tissue and graft material to be anastomosed
does not exceed 1.0 mm for the medium 1.4 mm clip applier, 1.6 mm for the large 2.0 mm clip
applier and 1.8 mm for the extra large 3.0 mm clip applier.
8. When using the small AnastoClip 0.9 mm clip applier on tissue, ensure that the total thickness of the
everted tissue does not exceed 0.5 mm.

Contraindications

1. Do not use the AnastoClip applier if tissue cannot be properly everted due to the presence of
arteriosclerotic or calcifi ed material, or where the vessel has been severely compromised due to
endarterectomy (e.g., carotid or any other artery in this condition).
2. The AnastoClips should not be more than 0.5 mm between one another. If this cannot be achieved,
C
Approximate
Clips Per
Overall Length
Applier
3.3 mm
35
D
Approximate
Clips Per
Overall Length
Applier
4.9 mm
25