8 Once all the OSUs are configured, you must launch a scan to detect
all the connected OSUs, clicking on the button F
9 Once scan is completed, the OSUs detection dialog box is
displayed with the OSU address, its serial number and C/P data.
Fig. 57
Scan results
10 Click on the Switch tab to apply the configuration (see
the optical switch" on page
OSU-9600 configuration with one OSU as multiplexer
1 Disconnect all the OSU-9600 that can be connected to the OTU
before performing the configuration.
2 Plug the OSU-9600 which will be used as multiplexer for the others
via the RS485, onto the OTU-8000
3 Enter the OSU address 31 for this OSU-9600
4 Click on A
PPLY
5 Once completed, a green text confirm the address has been
successfully applied
6 Press F
S
to check the configuration for this OSU multi-
ULL
CAN
plexer.
7 Once completed, unplug this OSU and plug the first OSU-9600,
which will be connected to OSU multiplexer, to the OTU-8000
8 Enter address 1 for this OTU and click on A
9 Once valid (see
Figure 56 on page
plug the second OSU-9600, which will have the address 2, to the
OTU-8000.
10 Enter address 2 on the OSU-9600 Settings page and click on
A
.
PPLY
11 Once valid, repeat steps
The addresses must be consecutive! (1, 2, 3, 4...).
12 Once all the OSU-9600 are configured, you must launch a scan to
detect all the connected OSU-9600, clicking on the button F
S
.
CAN
13 Once scan is completed, the OSU-9600 detection dialog box is
displayed with:
– the OSU-9600 Address 31, which represent the distributor OSU,
its serial number and C/P data
User Manual
7OTU80010/06
Chapter 3 Installation of the OTU-8000
OTU-8000 and WEB Interface
58).
PPLY
48), disconnect the OSU and
8
and 9.for the following OSUs.