Télécharger Imprimer la page

Civco Verza Mode D'emploi page 8

Bracelet de suivi avec système guidage

Publicité

English
3. Allow bracket to soak in disinfectant solution for at least the minimum contact time listed in disinfectant
manufacturer's instructions for use.
4. Rinse thoroughly by fully immersing bracket in large volume of utility water, agitate, and allow to set for at
least one minute.
5. Repeat Step 4 with new bath of utility water.
6. Remove bracket from utility water.
7. Dry with sterile, lint-free cloth. Visually inspect to ensure all surfaces are clean and dry.
HIGH-LEVEL DISINFECTION - CIDEX
NOTE: High-Level disinfection efficacy of this process was proven using CIDEX
1. Prepare high-level disinfectant according to manufacturer's instructions.
2. Fully immerse bracket in disinfectant solution. Use syringe to flush and remove air bubbles. Focus on cracks,
crevices, and hard-to-reach areas. Actuate all moveable parts.
3. Allow bracket to soak in disinfectant solution for at least the minimum contact time listed in disinfectant
manufacturer's instructions for use.
4. Rinse thoroughly by fully immersing bracket in large volume of utility water, agitate, and allow to set for at
least one minute.
5. Remove bracket from utility water.
6. Dry with sterile, lint-free cloth. Visually inspect to ensure all surfaces are clean and dry.
HIGH-LEVEL DISINFECTION - GIGASEPT
NOTE: High-Level disinfection efficacy of this process was proven using gigasept
1. Prepare high-level disinfectant according to manufacturer's instructions.
2. Fully immerse bracket in disinfectant solution. Use syringe to flush and remove air bubbles. Focus on cracks,
crevices, and hard-to-reach areas. Actuate all moveable parts.
3. Allow bracket to soak in disinfectant solution for at least the minimum contact time listed in disinfectant
manufacturer's instructions for use.
4. Thoroughly rinse bracket under running utility water for one minute. Remove air bubbles by agitating and
flushing with syringe.
5. Dry with sterile, lint-free cloth. Visually inspect to ensure all surfaces are clean and dry.
STORAGE CONDITIONS
Avoid storing product in areas of temperature extremes or in direct sunlight.
n
Store in a cool, dry place.
n
NOTE: Bracket is validated for 5-year product life.
NOTE: For questions or to order additional CIVCO products, please call +1 319-248-6757 or 1-800-445-
6741 or visit www.CIVCO.com.