PERFORMANCE CHARACTERISTICS
Tracking Brackets are designed to hold EMT sensors in close proximity to the ultrasound transducer while
n
connected to the ultrasound system.
The Verza Needle Guidance System's innovative design reduces blind zone and unguided needle length
n
while providing high stability for improved accuracy. Verza offers maximum versatility with a five-angle
approach for improved anatomical access while also featuring an expanded gauge range, accommodating
14 different sizes of interventional instruments. Verza features easy-to-read gauge sizes, angle indicator
and repeatable bolt action quick-release for secure instrument detachment.
The Verza Bracket design includes a simplified multi-angle design free of lock pins and removable parts.
n
The bracket also offers a low-profile fit, sleek design for ease of cleaning, and secure user-friendly closure
door for ease of application and removal.
Cover serves as a viral barrier to protect patients, users, and equipment from cross contamination.
n
NOTE: For a summary of clinical benefits for this product, visit www.CIVCO.com.
CAUTION
Federal (United States) law restricts this device to sale by or on the order of a physician.
WARNING
Verza Guidance System is designed and validated for use with transducer(s) listed on the cover of this
n
instruction.
Before use, you should be trained in ultrasonography. For instructions on the use of your transducer, see
n
your system's user guide.
Tracking bracket is for use with CIVCO general purpose sensors only.
n
General purpose sensor is for use with CIVCO accessories only.
n
Prior to use, ensure general purpose sensor is tracking properly.
n
Prior to use, refer to your system's user guide for use of general purpose sensor.
n
To avoid possible position distortion, keep general purpose sensor away from metal.
n
Review product labeling for sterile symbol. If sterile symbol appears on label, cover and guide are
n
sterilized using ethylene oxide.
Sterile product is intended for single-use only.
n
Do not use if integrity of packaging is violated.
n
Do not use if expiration date has passed.
n
Do not reuse, reprocess or resterilize single-use device. Reuse, reprocessing or resterilization may create a
n
risk of contamination of the device, cause patient infection or cross-infection.
CIVCO's guidance systems are designed and validated as an integrated two-part system consisting of a
n
custom reusable bracket and a disposable guide. Use of non-CIVCO guides on a CIVCO bracket or
locating feature may result in serious injury to the patient and/or user. To ensure safe and effective use in
compliance with system software guidelines, only CIVCO guides are recommended.
Bracket is packaged non-sterile and is reusable. To avoid possible patient contamination, ensure bracket
n
is properly cleaned, sterilized or disinfected before each use.
Do not use if bracket is damaged or does not fit properly.
n
Do not use if locating feature is damaged or guide does not fit properly.
n
Ensure gauge insert is in closed position prior to beginning procedure.
n
Use only water-soluble agents or gels. Petroleum or mineral oil based materials may harm cover.
n
For illustration purposes only, transducer and bracket may be shown without a transducer cover. Always
n
place a cover over transducer and bracket to protect patients and users from cross-contamination.
Ensure no air bubbles are present prior to use. Air left between cover and transducer lens may cause poor
n
image quality.
If the product malfunctions during use or is no longer able to achieve its intended use, stop using the
n
product and call CIVCO.
Report serious incidents related to the product to CIVCO and the competent authority in your Member
n
State or appropriate regulatory authorities.
NOTE: Product is not made with natural rubber latex.