BOLT-ACTION QUICK-RELEASE
1. Rotate quick-release tab to open guide insert funnel.
REMOVING GUIDANCE SYSTEM
1. Rotate guide locking lever to unlocked position to remove guide from VerzaLink locating feature of bracket.
2. Remove and dispose of single-use components according to facility infection control policies.
3. Lift bracket latch to remove from transducer. Retain bracket for reprocessing.
REMOVING GENERAL PURPOSE SENSOR FROM BRACKET
1. Rotate general purpose sensor out of bracket.
2. Pull general purpose sensor away from bracket to remove.
REMOVING BRACKET FROM TRANSDUCER
1. Pull bracket away from transducer to remove. Bracket may come off transducer forcefully. Remove with
caution.
REPROCESSING
WARNING
Users of this product have an obligation and responsibility to provide the highest degree of infection
n
control to patients, co-workers and themselves. To avoid cross-contamination, follow infection control
policies established by your facility.
Dispose of single-use components as infectious waste. Clean and sterilize reusable components after each
n
use.
See your ultrasound system's user guide for reprocessing transducer between use.
n
Refer to general purpose sensor instruction for reprocessing.
n
This procedure for reprocessing has been validated for effectiveness and compatibility. Device could be
n
damaged or cross-contaminated due to improper reprocessing.
Do not gas sterilize or autoclave bracket or general purpose sensor.
n
If bracket damage is evident, discontinue use and contact your CIVCO representative.
n
Inspect general purpose sensor for damage such as cracks or breakage. If damage is evident, discontinue
n
use of general purpose sensor and contact CIVCO.