Errors and mes-
sages
Relay supply
voltage fault
Internal sensor
fault
1) Fault LED of the module.
2) SFR and SSR of the Docking Station.
9.2

HART

The following errors and messages are only displayed if HART
components are present in the REGARD 3000 system.
Errors and mes-
sages
Accessories: flow
failure
Accessories: flow
warning
Device: tempera-
ture warning
Maintenance: cali-
bration incomplete
Maintenance: cali-
bration interval
exceeded
Optics: beam block
fault
Optics: beam block
warning
Optics: misalign-
ment
Power Supply: bat-
tery warning critical
|
Instructions for use
REGARD 3000
x
x
-
x
x
-
®
Cause
Pump gas flow too low, mea-
surement is no longer reli-
able
Pump gas flow too low, mea-
surement accuracy reduced
Sensor or device is not oper-
ating in the defined tempera-
ture range
Incorrect calibration
Calibration interval expired
The light transmission is so
weak that the gas concentra-
tion cannot be calculated
with the required accuracy
The device detects weak
light transmission, most likely
due to the contamination of
the optical system
Incorrect alignment of opti-
cal transmitter and receiver
Battery dead
Fault elimination and messages
Cause
Supply voltage of relay
outside approved range
Module error, internal
error with electronic
redundancy
Remedy
Check the hoses for block-
ages and adapt the pump
flow if necessary.
Check the hoses for block-
ages and adapt the pump
flow if necessary.
Operate the sensor or device
in the defined temperature
range.
Repeat calibration.
Calibrate transmitter.
Clean accessible optical ele-
ments.
Clean accessible optical ele-
ments.
Realign the optical transmit-
ter and receiver.
Immediately replace or,
where applicable, charge the
battery.
|
en
Remedy
Check and correct sup-
ply voltage. If error con-
tinues to occur, replace
module
Replace module.
®
-capable
169