3. In the module row, select the
4. Select Module maintenance.
5. Select the LED test: button.
 The LED test: button displays ON. The module LEDs flash.
6. Check that all module LEDs are working.
7. Select the LED test: button to exit the LED test.
 The LED test: button displays OFF. The module LEDs stop flashing.
The Module maintenance function allows a module to be restarted and is required
to delete a latching error on a module. A restart briefly shuts down the transmitter
powered by the module and thus restarts the transmitter.
Transmitters with an external power supply must be manually restarted by
disconnecting them from their power source. Take the warm-up time of the
transmitter into account.
– Alarms are suppressed during restart.
– All relays are briefly disconnected from the power supply in the case of an output
module. All channels configured as normally energized thus switch to active
during the restart (approx. 20 s).
To restart a module:
1. Select the
2. Select the Docking Station in which the module is installed.
3. In the module row, select the
4. Select Module maintenance.
5. Select Restart.
 A wizard is opened.
6. Follow the directions of the wizard.
8.4.3

Maintenance of channels and ports

The various functions for maintaining channels and Ports are described below. The
functions are identical for channels and Ports.
To display the configuration of channels or Ports:
● Channel: Select [Channel] > Configuration in the channel list view.
● Port: Select the
button in the channel list view.
✓ The configuration of the channel or Ports is displayed.
Observe the following information: "Configuration role", page 147.
To display the maintenance menu:
● Channel: Select [Channel] > Maintenance in the channel list view.
● Port: Select the
[Module] > [Port] > Maintenance.
✓ The maintenance menu for the channel or port is displayed.
|
Instructions for use
REGARD 3000
button.
button in the channel list view.
button.
> [Docking Station] > [Module] > [Port] > Configuration
button in the channel list view. Select [Docking Station] >
|