Maintenance

Errors and mes-
sages
Power Supply: bat-
tery warning low
Power Supply: volt-
age out of range
Sensor: failure
Sensor: vitality fair Approx. half of the sensor life
Sensor: vitality
poor
Sensor: warm-up
Sensor: warning
Unknown asset
10
Maintenance
REGARD 3000 is maintenance-free.
The alarm function (including the horn and light ring) and function of the SSR and
SFR must be checked during commissioning and then every 12 months.
The display function (screen test) must be checked when commissioning and then
every 12 months. For further information see: "Display Unit settings", page 147.
For further information see: "Maintaining the Docking Station", page 142.
For further information see: "Maintenance of channels and ports", page 145.
The check must be carried out by trained personnel. Attention must also be paid to
the maintenance intervals of the connected input and output elements.
10.1

Replacing module (same slot and same configuration)

For example, replace a faulty module with a new module of the same type in the
same slot.
A module can be replaced during operation.
1. Set all input and output channels to the Inhibit state.
2. Disable the alarm device connected to the SFR of the Docking Station.
3. Select the MUTE button to mute the horn.
4. Open configuration mode.
5. Save configuration.
170
Cause
Battery low
The power supply is outside
the defined range
Contacts or conductivity of
the sensor is too low for a
reliable measurement
time has been reached
Sensor life span expired or
the sensor has been over-
stressed
Sensor has not fully warmed-
up, a higher measuring toler-
ance is to be expected
Contacts or conductivity of
the sensor are close to an
error state
®
Connected HART
device
not fully supported
Instructions for use
Remedy
Battery needs to be replaced
or, where applicable,
charged soon.
Check and correct supply
voltage.
Replace or, where applica-
ble, top up sensor.
n/a
Sensor should be calibrated.
If the vitality remains low,
replace the sensor.
Wait until the sensor has fully
warmed-up.
The sensor should be
replaced or, where applica-
ble, topped up soon.
Update software.
|