10.7

Cleaning the Water Reservoirs and Reservoir Filters

Frequency: Monthly or every 100 cycles.
To avoid accidental cross-contamination, always start with the clean water reservoir and complete steps 1-6 BEFORE cleaning
the Venturi reservoir.
Follow the draining instructions in Section 10.6 to drain the reservoirs completely.
1.
Using a soft bristle brush, scrub the corners and loosen
any deposits.
Wipe the reservoir's surfaces using OPTIM wipes or a
2.
clean, lint-free cloth dampened with water.
Rinse the reservoir's surfaces with distilled water and
drain it using the front drain tube (See Section 10.6
Draining the Reservoirs for Cleaning).
3.
Remove the reservoir filter by pulling up on the filter's
tab.
Rinse the filter under running water and dry it before
4.
re-installing.
5.
Repeat these steps to clean the Venturi reservoir.
6.
Once you have cleaned both reservoirs, fill them with
distilled water and run one empty cycle.
The Venturi reservoir is connected to the back panel condenser. The water level in the Venturi reservoir may drop
as it tops up the condenser. Wait a few minutes and add more water to fill the Venturi reservoir.
CAUTION! To avoid cross contamination, be sure to use a different cloth and container with solution to wipe the
internal surfaces of each reservoir.