Emergency Door Opening - SciCan STATCLAVE G4 Manuel De L'utilisateur

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

6.4

Emergency Door Opening

The STATCLAVE is equipped with a safety mechanism that automatically regulates the chamber pressure when the unit loses
power. (The unit will take approximately 2 minutes to depressurize. Without power, the door lock will remain engaged.)
To unlock the door without power, follow these steps:
1.
Remove the emergency door unlocking pin located
in the handle of the tray extractor supplied with your
STATCLAVE
Insert the emergency door unlocking pin into the small
2.
hole on the side of the unit's handle. Push the pin into
the hole as far as it goes to trigger the door release.
3.
Remove the emergency door unlocking pin from the
hole in the door handle.
4.
Pull up on the handle to open.
CAUTION! Risk of Injury.
Do not force the door handle. If the unit is locked due to a cycle fault, do not force the door handle. Power OFF the
unit and allow it to cool for 10 minutes before attempting again.
32