Set Up; Installing Your Statclave - SciCan STATCLAVE G4 Manuel De L'utilisateur

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

2.

Set Up

2.1

Installing Your STATCLAVE

Place the STATCLAVE on a flat level surface strong enough
to support a 175 lbs (79.5 kg) load.
Allow for at least 7" (180 mm) of space ABOVE the unit
to enable access to the reservoirs. If there is less than
7" (180 mm) above the unit, use a screwdriver to remove
the front hinge of the clean water reservoir so that the
reservoir panel can slide in and out of position from the front
of the unit.
On the right side of the unit, allow for at least 2" (50 mm) of
space for ventilation.
The unit vents warm air from below the door. Ensure the
door overhangs the level surface.
Unit Dimensions and Operating Environment
Height with front legs retracted
Height with front legs fully extended
Width
Depth
Weight (empty)
Weight (with full reservoirs and full load)
IMPORTANT! To improve drainage, ensure the unit is tilted
toward the back. Use the leveling feet to ensure the front of
the unit is 1" (25 mm) from the level surface.
Temperature and Humidity
Avoid installing your STATCLAVE in direct sunlight or close to a heat source such as vents or radiators. The recommended
operating temperatures are between 5˚C-40˚C / 41˚F-104˚F with a maximum humidity of 80%.
Electromagnetic Environment
Your STATCLAVE has been tested and meets applicable standards for electromagnetic emissions. While your unit does not
emit any radiation, it may itself be affected by other equipment that does. We recommend that your unit be kept away from
potential sources of interference.
7"
19"-
19.5"
19" / 483 mm
19.5" / 495 mm
17.75" / 450 mm
25" / 635 mm
136 lbs/ 61.7 kg
175 lbs/ 79.5 kg
10
17.5"
1" gap
2"
25"