Using Rigid Sterilization Containers
The STATCLAVE is capable of processing re-usable rigid sterilization containers, which can be used as an alternative to
wrapping cassettes. These provide a convenient way to organize and store instruments, and cut down on the waste involved
with using paper wraps. Check the sterilization container manufacturer's instructions to determine its suitability for pre-vacuum
steam sterilization.
When Using Rigid Sterilization Containers:
DO ensure drying plates are used and that the rigid
sterilization containers are placed vertically in the rack
to promote proper drying. (See Inserting drying
plates, above.)
DO ensure there is space for sufficient air flow
between the rigid sterilization containers when loading
more than one container per row.
5.3

Unwrapped Instruments

Unwrapped instruments are also referred to as Immediate Use Instruments because once exposed to ambient or external
conditions, they cannot be maintained in a sterile state. If you plan to store your instruments after sterilization, wrap them
according to the instrument manufacturer's instructions, select the appropriate wrapped cycle and allow it to run to completion.
DO use the trays provided with your unit to hold
unwrapped instruments.
DO always use the trays with the chamber rack provided.
DO always use the tray extractor provided to remove trays
from the sterilization chamber.
DO arrange instruments made of different metals
(stainless steel, tempered steel, aluminum, etc.) on
different trays or keep them well separated from
each other.
DO arrange receptacles upside down to prevent water
from pooling inside.
DO ensure objects on trays are always arranged with
some distance between them ensuring they will remain
in the same position for the entire sterilization cycle.
DO ensure that hinged instruments are sterilized in
an open position.
DO position cutting instruments (scissors, scalpels, etc.)
so that they do not come into contact with each other
during sterilization.
DON'T load the trays beyond their maximum indicated
limit. (See STATCLAVE Load Capacities chart at the start
of Section 5. Loading Instruments)
DON'T stack trays or put them in direct contact with the
sterilization chamber walls.