SciCan STATCLAVE G4 Manuel De L'utilisateur page 58

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Problem
Time and date are
incorrect.
USB storage device does
not contain the last print
out.
Touchscreen shows
next
to the connectivity icon.
Unit's total cycle time is
too long.
Door will not open.
Door will not open – no
power.
Instruments are blackening
or there is damage to
materials.
Water remains in the
chamber at the end of a
cycle.
Possible Cause
Unit was shipped to a new time
zone.
USB device failure.
Unit is not connected to
Internet.
Unit is starting with a cold
chamber.
Chamber filters are clogged.
Lock is still engaged.
Power failure.
Sterilization temperature is too
high for the materials.
Chamber filters are obstructed.
Drying time is too short.
Possible Solution
The time and date are set on the date of manufacture but have not been
adjusted for a new time zone.
See Section 8. Using and Changing Settings.
Re-insert the USB storage device and wait for the data to copy over again.
If the problem persists, back up all the information you have on the USB
device and reformat it.
NOTE: You can always access all of your unit's cycle information through
the unit's web portal.
See Section 9. Storing, Retrieving and Printing Sterilization Records.
A red X over the connectivity icon means the unit is not connected to a
network. If the STATCLAVE is supposed to be connected to a network and
the X is visible, it is because the unit is unable to acquire an IP address.
To resolve the issue, try some of the following:
Check that the router is functioning properly
Check the LAN cable (try a new cable if possible)
Make sure your router assigns IP addresses automatically. Renew the IP
address by following these steps:
Scroll through the USER menu to NETWORK SETUP and select.
Select RENEW IP
Starting a cycle when the unit is cold can add as much as 12 minutes to the
total cycle time. Reduce the warm-up time between cycles or set the unit
to warm-up at a specific time in the morning. See Section 8.3 Setting the
Stand-by Mode.
Debris in the chamber filters can have an affect on cycle times. Remove the
chamber rack and inspect the chamber filters. Clean as necessary.
See Section 10.9 Cleaning the Chamber Filters.
Press the LOCK icon to go to the door lock status screen and press the
UNLOCK icon. See Section 1.7 Unlocking the Door.
This screen will show if the chamber is under pressure or vacuum and
whether it can be opened. See Section 6.2 Stopping a Cycle.
Use Emergency Door Unlocking procedure.
See Section 6.4 Emergency Door Opening.
The sterilization program selected is not appropriate for the materials/
instruments being sterilized. Check the materials/instruments
manufacturer's recommendations. Also see Section 7. Sterilization Cycles.
Inspect the three chamber filters and clean or replace them as needed.
See Section 10.9 Cleaning the Chamber Filters.
Drying times can be adjusted. See Section 8.2 Setting Drying Time.
56