Presetting your Bowie-Dick test
To schedule a Bowie-Dick test before the start of the next working day, use the delayed start function.
1.
At the end of the working day, select the Bowie-Dick
test.
2.
Press the DELAYED START icon.
3.
Enter the number of hours and minutes for the delayed
start.
Press the FORWARD icon to begin the countdown.
4.
Press the BACK icon to CANCEL.
The Bowie-Dick test will finish faster when the unit starts with a warm chamber. Set your test to start at least 15
minutes after your unit has begun to warm the chamber.
NOTE: The STATCLAVE's default setting is to keep the chamber temperature at Stand-by HIGH from 7:00 a.m. to
8:00 p.m.