Connecting Directly to a Drain
Any new central drain point installation should be done by a technician. The drain points (C) must be located on the upper
portion of vertical drain pipe ABOVE the P-trap.
For direct-to-drain connections, you will need to use the direct-to-drain hardware (provided with the unit).
1.
The unit is shipped with plugs in its ports. To remove a
plug, apply even pressure to the Inner Ring holding it
down on either side of the plug and pull out the plug.
Insert one elbow connector with silicone tube to the
2.
metal exhaust port (A).
3.
Insert the other elbow connector with silicone tube to
the plastic reservoir overflow
drain port (B).
4.
Connect the exhaust tube with elbow fittings (A) to the
port installed on the drain pipe (C).
5.
Connect the reservoir overflow tube (B) to the remaining
port on the drain pipe (C).
IMPORTANT! Avoid excess sagging in the lines; cut both tubes to measure.
IMPORTANT! Tubes should not be kinked, bent or otherwise obstructed. The connection point to the central drain
must be lower than the autoclave's support surface otherwise the reservoirs may not drain correctly.
12