The custom cycle's default cycle is 132°C / 270°F for 4 minutes with 25 minutes of drying. To adjust the settings of this
cycle, select SETTINGS then USER and follow these steps:
1.
From the Setup menu, scroll to Custom Cycle and
select.
From the Custom Cycle menu, press Cycle temperature.
2.
3.
Select one of the three sterilization temperature options.
Press the BACK icon to save your change.
4.
From the Custom Cycle menu, press Sterilization time.
IMPORTANT! Each temperature option comes pre-set with a minimum sterilization time and drying time. Follow the
instrument manufacturer's instrument reprocessing instructions when setting time and temperature. Failure to do so
could result in damage to the instruments and/or autoclave.
5.
Adjust how long the cycle will hold the sterilization
temperature. Press BACK to save your change.
Press Drying time to adjust the amount of drying. Press
6.
BACK to save your change, then press BACK again to
exit the Custom Cycle settings.
7.
To confirm your changes, select the Custom Cycle icon
from the home screen and check the cycle description
at the top of the start screen. If it is incorrect, return to
the Custom Cycle settings and re-enter your selections.
36