Error Code

CF 4
CF 10
CF 16
CF 17
CF 43
CF 44
CF 73
CF 77
Possible Cause
Door seal leak.
Blocked exhaust.
Steam generator is too hot.
Initial vacuum phase
unsuccessful.
Problem with the automatic
clean water filling system.
Venturi reservoir is too full.
Unable to reach vacuum.
Unable to reach vacuum.
What you can check before calling for service
Check door seal for obstructions.
Check that the door seal is properly seated in the door channel.
Check the exhaust tubing at the back of the unit to make sure it is not blocked
or kinked.
Check that the exhaust tubing does not have a deep sag that could impede
the flow of exhaust water.
Check that the water levels in both reservoirs are adequate.
(See Section 4 Filling the Water Reservoirs).
Check door seal for obstructions.
Check that the door seal is properly seated in the door channel.
Check the auxiliary pump to see that it is correctly connected.
(See Section 4.4 Automatic Filling Using External Water Tank and Auxiliary
Pump).
Check if the unit is set to automatic filling but there is no pump connected.
(See Section 4 Filling the Water Reservoirs).
Check for a blocked or kinked filling tube.
Check for a blocked exhaust tube.
Check the Venturi reservoir overflow sensor (the float sensors on the back
wall of the reservoir) to make sure it is not simply stuck in the high position.
Sometimes the overflow sensor is unintentionally activated during a reservoir
cleaning. This can prompt a CF 44 message. Press the STOP icon to reset.
Check door seal for obstructions.
Check that the door seal is properly seated in the door channel.
Check door seal for obstructions.
Check that the door seal is properly seated in the door channel.
58