Manual Filling Using Quick Connector At Unit Front - SciCan STATCLAVE G4 Manuel De L'utilisateur

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

4.2

Manual Filling Using Quick Connector at Unit Front

The front fill quick connector is for the refilling of the clean water reservoir. The Venturi reservoir does not have a front fill quick
connector because it is only filled on installation or after cleaning.
1.
To use this method, you will need the manual fill
container (0.5 US gal / 2 L) with quick connect and
tubing (sold as an accessory).
2.
Fill the accessory container with distilled water, keeping
it horizontal.
3.
With the unit powered ON, connect the tube's quick
connector to the unit's clean water fill port at the front.
4.
Hold or place the manual fill container at a level higher
than the unit's fill port and loosen the cap on the
container to allow water to flow into the tank.
5.
With the door open, access the LCD's home screen
and select any cycle to view the pre-cycle information
screen.
6.
The clean water level check mark will turn yellow when
the minimum water level is reached.
You can start a cycle if water level icons are either yellow or green.
Repeat this procedure a second time to fill the water
7.
reservoir to max level. (Each fill takes 4 minutes.)
Continue until the clean water level indicator turns green
and you hear three BEEPs.
If your unit is not directly connected to the drain, empty the waste bottle regularly or each time you are filling the
clean water reservoir.