Connecting the Waste Bottle
The unit is shipped with a waste bottle in case a direct-to-drain installation is not possible. To use the waste bottle, follow these
steps:
Set the waste bottle (D) on the ground or in the
1.
cabinetry below the unit.
2.
Connect the exhaust tube (A) to a port on the waste
bottle cap (C).
3.
Connect the reservoir overflow tube (B) to a port on the
waste bottle cap (C).
IMPORTANT! Avoid excess sagging in the lines; cut both
tubes to measure.
IMPORTANT! Tubes should not be kinked, bent or
otherwise obstructed. The waste bottle must be lower than
the autoclave's support surface otherwise the reservoirs
may not drain correctly.
Add a small amount of water to the empty waste bottle to
give it stability.
2.4

Connecting your STATCLAVE to a Network

Connecting to a Wired Network
Do you want to drain the unit prior to shipping or cleaning? For instructions on how to completely drain both
reservoirs for cleaning or shipping, see Sections 10.6 and 10.14.
1.
Connect Ethernet cable to port at back of unit.
2.
From the home screen select the CONNECTIVITY icon.
3.
Confirm that the unit is connected to both the Network
and Internet and press the back icon to return to
the home screen.
13