Running a Bowie-Dick Test

The BOWIE-DICK test is used to ensure proper air removal is occurring in a pre-vacuum autoclave. Complete air removal is
important because pockets of cool air remaining in the chamber can compromise sterilization. The Bowie-Dick test runs a
cycle at 134°C / 273°F for 3.5 minutes to evaluate the correct air removal.
To perform a Bowie-Dick test, you will need a Bowie-Dick device or test pack. These are NOT supplied with your STATCLAVE.
To perform the test, follow the instructions provided by the test pack manufacturer.
Generally, the process is as follows:
Open the unit's chamber door to insert a Bowie Dick
1.
test pack. (Position pack at centre back.)
2.
Close and lock the door.
From the home screen, select the TESTS icon.
3.
4.
To run a Bowie Dick Test, press the BD icon.
Press the START icon.
5.
Allow the test to run to completion.
6.
Press the STOP icon to release the load.
Follow the test manufacturer's instructions to interpret
the test results.
If the unit passes the test, it is ready for use. If the unit fails,
check the test manufacturer's instructions and repeat the test.