Cleaning The Door Seal And Door Plate; Cleaning And Disinfecting The External Surfaces; Cleaning The Sterilization Chamber, Rack And Trays - SciCan STATCLAVE G4 Manuel De L'utilisateur

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

10.3

Cleaning the Door Seal and Door Plate

Frequency: Daily
CAUTION! Hot Chamber.
The unit's Stand-by mode is preset to maintain a hot
chamber. Turn the unit off and allow adequate time for it to
cool before performing maintenance.
Using a clean, lint-free cloth dampened with water, wipe
1.
the door seal and door plate.
10.4

Cleaning and Disinfecting the External Surfaces

Frequency: Clean daily. Disinfect weekly.
Clean all of the STATCLAVE's external parts using
1.
OPTIM wipes or a clean, lint-free cloth dampened with
water and, if needed, a mild detergent.
2.
Dry the surfaces and remove any residue before
powering ON the unit.
10.5

Cleaning the Sterilization Chamber, Rack and Trays

IMPORTANT! When cleaning the chamber, be careful not to damage the temperature probe on the inside back wall
of the chamber.
CAUTION! Hot Chamber.
The unit's Stand-by mode is preset to maintain a hot chamber. Turn the unit off and allow adequate time for it to cool
before performing maintenance.
Frequency: Monthly or every 100 cycles
1.
Remove the sterilization trays and the rack from the
chamber. (Pull the rack out to disengage it).
2.
Use a clean, lint-free cloth dampened with water to
clean the chamber and the chamber flange. Wipe dry.
3.
Use a clean, lint-free cloth dampened with water to
clean the STATCLAVE rack and trays.
49