Operating faults

12.1

Metering warning messages (display)

Display
reserve report (flashing)
Fig. 12.1 Operating faults: Metering warning messages (display): reserve report
empty report
Fig. 12.2 Operating faults: Metering warning messages (display):empty report
metering lock (only possible
if this is configured)
Fig. 12.3 Operating faults: Metering warning messages (display):metering lock
Indicator in operation mode
4..20mA flashes
standard signal monitoring
responds
Fig. 12.4 Operating faults: Metering warning messages (display):indicator in operation mode
standard signal is above
--.-
23.0 mA
Fig. 12.5 Operating faults: Metering warning messages (display):standard signal
12.2

Alarm messages (display)

NOTE
Appearing alarm messages can be quit by pressing the Start/Stop button.
Display
failure 1
permanent stroke
failure 2
no stroke
failure 3
motor control
failure 4
meter. monitor
failure 9
DongleBox
not found
failure 10
OGM PLUS
not found
failure 11
microflow
not found
failure 12
Communication
DongleBox
failure 13
Communication
microflow
417101438_EMP_II_E60
Meaning
Effect
pump continues
running
pump is stopped
pump is stopped
pump is stopped
pump runs in
continuous operation
Meaning
Motor runs
uncontrolled in
over dosage
continuous operation
Fig. 12.6 Operating faults: Alarm messages (display) failure 1
Motor is not running
despite of rotating
no metering
dosing symbol.
Fig. 12.7
Operating faults:Alarm messages (display) failure 2
motor in continuous
continuous
operation without
metering
request
Fig. 12.8
Operating faults:Alarm messages (display) failure 3
error in evaluating the
pump meters
metering monitoring
too little or too
or oval gear meter
much
Fig. 12.9
Operating faults:Alarm messages (display) failure 4
The dongle box is not
detected the next time
Pump is being
the pump is switched
stopped
on
Fig. 12.10 Operating faults:Alarm messages (display) failure 9
The OGM
Plus
is not
detected the next time
Pump is being
the pump is switched
stopped
on
Fig. 12.11 Operating faults:Alarm messages (display) failure 10
The MicroFlow is not
detected the next time
Pump is being
the pump is switched
stopped
on
Fig. 12.12 Operating faults:Alarm messages (display) failure 11
No comunication
between DongleBox
Pump is being
Box and pump
stopped
Fig. 12.13 Operating faults:Alarm messages (display) failure 12
No comunication
between MicroFlow
Pump is being
Box and pump
stopped
Fig. 12.14 Operating faults:Alarm messages (display) failure 13
Cause
low-level advance
warning active
empty report active
no external enable of
the pump
standard signal is
under 3mA or cable to
standard signal
connection is broken
standard signal
exceeds the display
range
Effect
Cause
power electronic failure
backpressure too high
valve closed at pressure
side
Motor overheated/damaged cool of the motor, or. change
power electronic failure
optical sensor polluted
PC-board fault
Plug connection from gear
to PC-Board interrupted
tube fault
diaphragm fault
backpressure too high or
too low
Connection between
DongleBox and pump
interrupted
Connection between OGM
PLUS and DongleBox
interrupted
Connection between
MicroFlow Box and pump
interrupted
Transmission of data from
the dongle box to the pump
has been disrupted during
operation
Transmission of data from
the MicroFlow to the pump
has been disrupted during
operation
- 128 -
Remedy
Refill the metering medium
Refill the metering medium
Activate external enable or deactivate
the metering lock in the configuration
menu. (see chapter 10.3.7)
check the standard signal or cable
reduce the standard signal
Remedy
replace PC-Board
reduce pressure
open valve
replace PC-board
send in pump
control optical sensor
connection
check the tubes
check the diaphragm
check the backpressure
Restore the connection or
disable the dongle box in the
software (see chapter 12.2.1)
Restore the connection or
disable the dongle box in the
software (see chapter 12.2.1)
Restore the connection or
disable the dongle box in the
software (see chapter 12.2.1)
Check the connection and
replace the dongle box if
necessary
Check the connection and
replace the MicroFlow if
necessary