10.3.13.2 Metering monitoring / Stop pump

10.3.13.3 Metering monitoring / Strokes

10.3.13.4 Metering monitoring / Deviation (only at
A tolerance in % is set under "deviation".
The actual incoming pulse rate recorded by the oval gear meter within the pump interval to be
monitored (see chapter
rate defined during calibration.
If this comparison results in a deviation that is greater than the % value configured under
"deviation", a "metering monitoring alarm" is triggered.
NOTE
417101438_EMP_II_E60
10.3.13.3
This function can only be selected if a dongle box is connected and configuration/oval
gear meter (see chapter 10.3.11) has been activated.
E60
PLUS
&
"Configuration/Strokes") is compared with the target pulse
- 112 -
Fig. 10.35 Metering monitoring / Stop pump
The "stop pump" function is used
to set whether or not the pump
should be stopped if a "metering
monitoring alarm" is triggered.
- Pump is not stopped when a
"metering monitoring alarm"
is triggered. (Default setting)
✓ Pump is stopped
when a "metering monitoring alarm"
is triggered.
Fig. 10.36 Metering monitoring / Strokes
The "strokes" option indicates the
stroke interval to be monitored.
Setting range: 0 – 100
Default setting = 10
OGM
or
OGM
PLUS
)
Fig. 10.37 Metering monitoring / Deviation (only at E60
PLUS
OGM or OGM
)
Setting range: 0-100%
Default setting = 20 %
PLUS
&