10.3.16.2
Degas
PLUS
10.3.16.3
Degas
PLUS
The time between two degassing intervals is set under "degas break".
If a metering request is submitted during this period, this is processed and the
break time is reset and restarted.
Setting range:
10.4
Calibration
10.4.1
Overview
The procedure and display during the calibration process differ depending on whether or not
an oval gear meter is connected.
Fig. 10.54 Calibration Overview
417101438_EMP_II_E60
/ Degas strokes
/ Degas break
hours
minutes
total
Fig. 10.52 Configuration Degas
The number of strokes that
should be processed during a
degassing interval is recoded
under "degas strokes".
Setting range: 1-100
Default setting = 0
00 - 23
00 - 59
00.01 - 23.59h
Calibration pump:
• see chapter
Calibration pump with
oval gear meter:
• see chapter
- 119 -
PLUS
/ Degas strokes
Degas
PLUS
Fig. 10.53 Configuration Degas
PLUS
/ Degas break
Default setting
= 00:00
10.4.2
10.4.3
Rev. 12-01.19