At the completion of the ablation cycle, a "Procedure Complete" screen
will appear with a summary of the procedure. The "Procedure Complete"
screen will capture the following information for each procedure:
• Cavity Length
• Cavity Width
• Power Level
• RF Ablation Time
2.39 Unlock the disposable device by pressing the lock release button.
Close the disposable device by holding the front handle stationary
and gently pulling the rear handle backwards until the closed array
indicator, located at the hinge of the front and rear handles, reads
"ARRAY CLOSED". This indicates that the array has been retracted
into the sheath and the disposable device is in the closed position.
NOTE: If it is difficult to close and remove the disposable device,
see the Troubleshooting section, "Difficulty closing and removing
the disposable device post-ablation".
CAUTION: To avoid damaging the device, employ gentle technique
when retracting the array.
2.40 Withdraw the disposable device from the patient.
2.41 TURN OFF THE NOVASURE RF CONTROLLER.
2.42 Perform postoperative patient care according to standard
procedures. The used disposable device must be treated as
biohazardous waste and disposed of according to standard
practices of the hospital or clinic where the treatment is performed.
2.43 Discharge the patient from the hospital or office as indicated by the
managing physician.
Periodic Maintenance and Service
There is no service manual for the NovaSure RF Controller since there
are no field serviceable components within the unit.
WARNING: No modification of this equipment is allowed.

Periodic maintenance

The RF controller has been designed and tested to meet IEC
60601-1 and other safety standards. Maintenance is not required
as the system performs self-checks when power is turned on.To
clean the controller refer to the "Cleaning and Sanitizing" section.
RF power output test
The NovaSure RF controller Model 10 integrates automatic power
output testing in a power on self test (POST). During the POST the
controller's power output (Pc) is delivered into a shunt resistor (Rs)
located inside the controller. Pc is targeted to be 180 watts and Rs
is nominally 25 ohms. During the POST, no power is delivered to
the disposable device connector at the front of the controller.
The NovaSure RF controller Model 10 performs POST
automatically upon power up. To access POST information, go to
Settings and scroll to System Information. Press the "View Info."
button.
NOTE: If a system fault occurs during the POST, toggle off the
power at the power input module and repeat the POST. If a
system fault occurs a second time, remove the RF controller
from service and contact Hologic customer service.
CAUTION: Do not attempt to repair the controller if problems
are suspected. Follow the troubleshooting guide in this
manual. If problems persist, call Hologic Technical Support for
instructions.
Sterile NovaSure disposable device: No maintenance is necessary.
Single-patient use only. Do not reuse or re-sterilize the NovaSure
disposable device.

NovaSure Model 10 RF Controller Screen Icons

The following is a description of the icons on the NovaSure Model 10 RF
controller screen.
Settings Icon:
Press this button to display the
setting options for Language,
Brightness, and Volume.
21
Last Procedure Icon:
Press this button to display the
"Procedure Complete" Screen
and review the summary from
the last procedure.
Help Icon:
Press this button to display more information
regarding the current step of the procedure.
This icon will display an abridged version of
the IFU language.
ENGLISH

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Nova sure 09 rfNova sure 10 rf

Table des Matières