7.2.3.5 Installing the metering monitoring

NOTICE!
The metering monitoring function must be activated in menu first. The selection
must also be made of whether the pins 2+4 (input "pulse") or 3+4 (input
"metering lock") should be used for the metering monitoring input
Ä Chapter 10.3.13 'Configuration / Metering monitoring' on page 82).
(
Fig. 17: Installing the metering monitoring
2 + 4 = Input metering monitoring function with potential-free contact with selection
input „pulse" at configuration / metering monitoring (
10.3.13 'Configuration / Metering monitoring' on page 82).
PIN
Assignment
2
Relay input
4
GND (┴)
1 +2 + 4 = Input metering monitoring function with electronic switch (NPN) with
selection input „pulse" at configuration / metering monitoring (
10.3.13 'Configuration / Metering monitoring' on page 82).
PIN
Assignment
1
5 V, DC
2
Relay input
4
GND (┴)
or
3 + 4 = Input metering monitoring function with potential-free contact with selection
input „metering lock" at configuration / metering monitoring (
10.3.13 'Configuration / Metering monitoring' on page 82).
PIN
Assignment
3
Relay input
4
GND (┴)
1 +3 + 4 =Input metering monitoring function with electronic switch (NPN) with
selection input „metering lock" at configuration / metering monitoring (
10.3.13 'Configuration / Metering monitoring' on page 82).
PIN
Assignment
1
5 V, DC
3
Relay input
4
GND (┴)
Ä Chapter
43
Device installation
Ä Chapter
Ä Chapter
Ä Chapter
417101442 Rev. 12-02.2019