Resistance to electromagnetic interference (immunity) for all devices and systems
The device is designed for use in electromagnetic environments specified below� The customer or op-
erator of the device should ensure that the device is operated such an environment�
Interference immu-
nity tests
Electrostatic dis-
charge (ESD) in acc�
with IEC 61000-4-2
Electrical fast tran-
sient/burst immunity
test in accordance
with IEC 61000-4-4
Voltage surge in ac-
cordance with
IEC 61000-4-5
Voltage drops,
short-term interrup-
tions and fluctua-
tions of the supply
voltage in accord-
ance with
IEC 61000-4-11
Magnetic field for a
supply frequency
(50/60 Hz) in ac-
cordance with
IEC 61000-4-8
Table 1: Resistance to electromagnetic interference (immunity) for all devices and systems
2141100009L30 1612V001
IEC 60601 - test
level
±6 kV contact dis-
charge
±8 kV air discharge
±2 kV for mains ca-
bles
±1 kV for input and
output cables
±1  kV voltage outer
conductor/outer
conductor
±2  kV voltage outer
conductor/earth
< 5% U
(> 95%
T
drop in U
) for 1/2
T
period
40% U
(60% drop
T
in U
) for 5 periods
T
70% U
(30% drop
T
in U
) for 25 periods
T
< 5% U
(> 95%
T
drop in U
) for 5 s
T
3 A/m
Compliance level
±6 kV contact dis-
charge
±8 kV air discharge
±2 kV for mains ca-
bles
±1 kV for input and
output cables
±1  kV push-pull
The quality of the supply voltage
voltage
should correspond to a typical
commercial or hospital environ-
±2  kV common
ment�
mode voltage
< 5% U
(> 95%
T
drop in U
) for 1/2
T
period
40% U
(60% drop
T
in U
) for 5 periods
T
70% U
(30% drop
T
in U
) for 25 periods
T
< 5% U
(> 95%
T
drop in U
) for 5 s
T
3 A/m
Appendix
Electromagnetic environment
- guidelines
Floors should be made of wood
or cement, or covered with ce-
ramic tiles� If the floor is covered
by synthetic material, then the
relative humidity must be at
least 30%�
The quality of the supply voltage
should correspond to a typical
commercial or hospital environ-
ment�
The quality of the supply voltage
should correspond to a typical
commercial or hospital environ-
ment� If the operator of the de-
vice needs the unit to continue
working even if the mains power
supply is interrupted, we rec-
ommend powering the device
from an uninterruptible power
supply (UPS) or from a battery�
The magnetic fields at mains
frequency should be within the
range of typical values encoun-
tered in a commercial or hospi-
tal environment�