Molars
Bitewing
Check and adjust the specific X-ray unit in accordance with the standard values�
i

8.6 Acceptance tests

The required tests (e�g� acceptance tests) must be carried out in accordance with local rules and reg-
ulations�
Find out which tests are required�
i
Carry out testing in accordance with local rules and regulations�
i
Acceptance test
The Intra / Extra Digital test body is required for acceptance tests with the image plate and
sensor as receivers, and possibly also the corresponding test body holder�
Before the unit is started up and used for the first time, the acceptance test of the X-ray system
i
must be carried out in accordance with national regulations�
The constancy tests, which must be carried out at regular intervals by the surgery personnel, are
based on the results of the acceptance test�
Electrical safety checks
Carry out the electrical safety check according to the national law (e� g� in accordance with IEC
i
62353)�
Document the results�
i
2141100009L30 1612V001
Without X-ray field
limitation
70 kV
mGycm
2
0�14 s
12�8
0�14 s
12�8
DC emitter, 6  mA
Tube length 30 cm
X-ray field limitation
2x3
70 kV
mGycm
2
0�14 s
2�7
0�14 s
2�7
Installation
X-ray field limitation
3x4
70 kV
mGycm
2
0�14 s
5�4
0�14 s
5�4
23
EN