12 Maintenance

12.1 Recommended maintenance schedule

Only trained specialists or personnel trained by Dürr Dental may service the unit�
Prior to working on the device or in case of danger, disconnect it from the mains (e� g� pull the
mains plug)�
The recommended maintenance intervals are based on using the unit for 15 intraoral images per day
and 220 working days per year�
Maintenance
Maintenance work
interval
Annually
i
i
i
i
i
Every 3 years
i
i
i
2141100009L30 1612V001
Visually inspect the unit�
Check the image plates for signs of scratches and change if necessary�
Check the belt drives, transport belts and springs, and replace if necessary�
Remove dust and dirt from accessible parts�
Carry out a system check�
Change the wiper blade set�
Change the roller fixtures�
Change the drive belt�