Installation
Installing the unit with the wall mounting
bracket
The unit can be mounted on a wall with the wall
mounting bracket (see "3�3 Special accesso-
ries")�
For installation refer to the installation in-
structions for the wall mounting (order
no� 9000-618-162)

7.3 Electrical connections

Electrical safety when making connections
The unit must only be connected to a correct-
i
EN
ly installed power outlet�
Do not place non-fixed multi-socket units on
i
the floor� Follow the requirements in sec-
tion 16 of IEC 60601-1 (EN 60601-1)�
Do not operate any other systems using the
i
same multiple socket�
Make sure that none of the electrical cables
i
leading to the unit are under any mechanical
tension�
Before initial start-up check that the mains
i
supply voltage and the voltage stated on the
type plate match (see also "4� Technical
data")�
Connecting the unit to the mains
The unit has no main power switch� For
this reason it is important that the unit is
be set up in such a way that the plug can
be easily accessed and unplugged if re-
quired�
Requirements:
– Properly installed power outlet close to the
unit (observe the max� mains cable length)
– Easily accessible power outlet
– Mains voltage must match the information
shown on the type plate of the power supply
unit
Attach the matching country-specific adapter
i
to the power supply unit�
16
Remove the cover from the rear of the unit�
i
Plug in the connecting plug of the power sup-
i
ply unit into the connection socket of the unit�
Secure the cable with a cable clip�
i
2
Plug the mains plug into the power outlet�
i
Refit the cover�
i
The cover on the rear must be cor-
rectly fitted when the unit is operated
within the vicinity of the patients�

7.4 Connecting the unit

The device can be connected either via the USB
port or via the network connection� The unit can
only be operated via the network when using
VistaSoft/VistaConnect� The cables are included
in the scope of delivery�
Do not connect the device via the USB
port and via the network connection at
the same time�
If the device is connected via the USB
port and via the network connection at
the same time, the network connection
will take priority�
1