Combining devices safely
Take care when connecting units together or to
parts of other systems as there is always an ele-
ment of risk (e�g� due to leakage currents)�
Only connect units when there can be no
i
question of danger to operator or to patient�
Only connect units when it is safe to do so
i
and there is no risk of damage or harm to the
surroundings�
If it is not 100% clear from the unit data sheet
i
that such connections can be safely made or
if you are in any doubt, always get a suitably
qualified person (e�g� the manufacturer) to ver-
ify that the setup is safe�
Safety and essential performance features are
i
independent of the network� The unit is de-
signed for operation independent of a net-
work� However, some of the functions are not
available in this case�
Incorrect manual configuration can lead to
i
significant network problems� The expert
knowledge of a network administrator is re-
quired for configuration�
The data connection utilizes part of the band-
i
width of the network� Interactions with other
medical devices cannot be completely exclud-
ed� Apply the IEC 80001-1 standard for risk
assessment�
The unit is not suitable for direct connection to
i
the public internet�
Always comply with the relevant requirements
i
from of IEC 60601-1-1 (EN 60601-1-1) when
connecting the unit to other appliances, e�g�
to a computer system, both inside and out-
side of the patient environment�
Only connect peripheral units (e�g� computer,
i
monitor, printer) that conform at least to the
requirements set out in IEC 60950-1
(EN 60950-1)�
A copy of the system manufacturer's
declaration in accordance with Article 12
of Directive 93/42/EEC can be found in
our download section at www�duerrden-
tal�com (document no� 9000-461-264)�
2141100009L30 1612V001
Connecting the unit via the network cable
Purpose of the network connection
The network connection is used to exchange in-
formation or control signals between the unit
and a software installed on a computer, in order
to, e� g�:
– Display parameters
– Select operating modes
– Indicate messages and error situations
– Change unit settings
– Activate test functions
– Transmit data for archiving
– Provide documents concerning the units
Remove the cover from the rear of the unit�
i
Connect the supplied network cable to the
i
network connection of the device�
Refit the cover�
i
The cover on the rear must be cor-
rectly fitted when the unit is operated
within the vicinity of the patients�