Installation

Only qualified specialists or employees
trained by Dürr Dental are permitted to
install, connect and start using the unit�
6

Requirements

6.1 Installation/setup room

The room chosen for set up should fulfil the
following requirements:
– Closed, dry, well-ventilated room
– It should not be a room made for another
purpose (e�g� boiler room or wet cell)�
– Max� light intensity 1000  Lux, no direct sun-
light at the place of installation of the unit
– There should be no large fields of interference
(e�g� strong magnetic fields) present that can
interfere with the correct operation of the unit�
– Refer to the requirements for environmental
conditions in "4 Technical data"�

6.2 System requirements

The system requirements for the comput-
er systems can be found in the download
area at www�duerrdental�com (document
no� 9000-618-148)�

6.3 Monitor

The monitor must comply with the requirements
for digital X-ray with a high light intensity and
wide contrast range�
Strong ambient light, sunlight falling directly onto
the monitor and reflections can make it harder
or even impossible to perform a diagnosis
based on the X-ray images�
2141100009L30 1612V001
7
Installation

7.1 Carrying the unit

NOTICE
Risk of damage to sensitive compo-
nents in the unit as a result of shocks
or vibrations
Do not expose the unit to any strong
i
vibrations or shocks�
Do not move the unit during opera-
i
tion�

7.2 Setting up the unit

Portable and mobile HF communication appli-
ances can interfere with the effectiveness of
electrical medical devices�
Do not stack the unit next to or together with
i
other appliances�
If, however, this unit is operated next to other
i
units or stacked with other units, monitor the
unit carefully in the configuration selected in
order to ensure normal operation�
The unit can be set up as a tabletop unit or
mounted on a wall using the wall bracket�
The load-bearing capacity of the table or wall
must be suitable for the weight of the unit (see
"4 Technical data")�
Setting the unit on a table
To prevent errors when scanning the im-
age data, install the unit so it is not ex-
posed to vibrations�
Place the unit on a firm, horizontal surface�
i
EN
6,5
kg
15