10.4.4

Calibration / Manual

If the calibration value to be entered is known, a "dry calibration" (immediate entry of the
value without a previous calibration process) can be conducted.
However, this method is not very accurate as no consideration is given to the on-site
conditions (back-pressure, viscosity, cable cross-chapters and lengths etc).
Fig. 10.50 Calibration / Manual

10.4.4.1 Calibration data table

The pump capacity in l/h is allocated to the pump's electronics on the basis of the calibration
values entered in the table.
Example:
NOTE
These values relate to the metering medium of water at 20°C
Pump
Fig. 10.51 Calibration data table
10.5

Operating data

The following operating data is recorded and displayed under this menu item:
Operating hours
Liter
Amount of plulses
417101441_EMP_III_E60.docx
A specified value of 267 ml results in a pump capacity of 16 l/h.
Pump capacity [l/h]
16
25
54
80
120
Pump calibration value [ml] or [l]
- 117 -
267 [ml]
417 [ml]
900 [ml]
1,33 [l]
2,00 [l]
Rev. 13-10.19