Configuration / Autostart

This function determines whether the pump is set to 'OFF' ("Standby" in V 3.5) when the
mains connection is re-established following a power cut or if the pump should immediately
recommence functioning in the configured operation mode.

10.3.6.1 Selecting

10.3.7

Configuration / Metering lock

If the metering lock is enabled, the pump only functions if an external enable contact is
connected to plug II (or at plug VII or plug VIII, if a
and an external enable contact is connected (electrical connection see chapter
"Installing the control via the metering lock",
"Installing the metering lock"). The selected operating mode does not affect the use of the
metering lock.
10.3.7.1 Selecting
10.3.8

Configuration / Low-level contact

This function specifies whether an open or a closed contact at the level input (plug I,
see chapter
7.2.3.1
report") is regarded as level OK.
417101441_EMP_III_E60.docx
"Installing the suction pipe with low-level advance warning and empty
Fig. 10.20 Configuration / Autostart: Selecting
-
autist inactive.
When the mains voltage is applied, the
pump always enters the "OFF"
("Standby" in V 3.5) operating state.
(Default setting)
autostart ist active.
When the mains voltage is connected,
the pump starts functioning in the
configured operation state.
dongle box
is used) between PIN 3+
8.2.2.2
"Installing the metering lock" and
Fig. 10.21 Configuration / Metering lock: Selecting
-
interrogation of the metering lock
is inactive. (Default setting)
interrogation of the metering lock
is active.