3.1 Generator - Front View

POWER STATUS
PULSE COUNT
BALLOON
SIZE
THERAPY STATUS
CONNECTOR DOOR
CHARGER
CONNECTOR
(behind door)

3.2 Control and Indicator Functions

CONTROL
DESCRIPTION
POWER ON/OFF
Turns Generator on or off.
THERAPY
Press to activate the Generator.
ON/OFF
INDICATOR
DESCRIPTION
POWER STATUS
On green when Generator is on.
Indicator
On yellow when user action is required regarding the Lithoplasty
Catheter (CATH).
On red when internal diagnostics have detected a problem (SYS).
BATTERY
Indicates battery charge remaining.
CAPACITY
Display/
Charging Status
BATTERY
Lightning bolt symbol appears when the Charger Module is
CHARGING
connected and is charging the battery from mains power.
Indicator
BALLOON SIZE
Catheter Balloon Diameter and Length
Display
PULSE COUNT
Number of Pulses Delivered.
Display
THERAPY
On green when the device is ready to deliver therapy. Flashes to
LBL 60780 rev. B
POWER ON/OFF
BATTERY CAPACITY
THERAPY ON/OFF
THERAPY
CONNECTOR
MORE INFO
Refer to 3.1 Generator Front View
Connector Cable and a valid
Lithoplasty Catheter must be
connected to activate.
MORE INFO
Refer to 3.1 Generator Front View
See Section 4.0 step 6.
See Section 7.0.
See Section 4.0 step 2.
Charge Generator before use. See
Sections 5.2 and 5.3.
When Connector Cable and valid
catheter connected.
Increments during treatment as each
pulse is delivered
See Section 4.0, Steps 5 – 9.