5.2 Connecting to Line Power

The Generator normally operates from an internal, rechargeable battery system. However, in order to charge
the battery system, the Charger Module must first be connected to line power. Locate the AC Mains Cable
shipped with the Generator and connect it to the AC Mains input as indicated in the figure on the next page.
Choose a safe location where connection of this power cable to AC Mains will not create a trip hazard and
connect it to the AC Mains outlet.
The Charger Module is designed to operate from power supplies world-wide. See Appendix C for more
information.
WARNING
SHOCK HAZARD
To avoid the risk of electrical shock, this equipment must only be connected to a grounded electrical outlet (electrical supply
mains with protective earth.) Use with a hospital grade receptacle. Grounding reliability can only be achieved when connected
to an equivalent receptacle marked "hospital use" or "hospital grade". Only use the Charger Module provided with the
Generator to avoid shock.

5.3 Charging the Internal Battery

Charging the battery requires the Charger Module be connected to AC Mains, and the Charger Module must
also be connected to the CHARGER CONNECTOR located on the front panel of the Generator (see Section 3.1).
It will be necessary to disconnect the Connector Cable from Generator, if one is attached. To do so, gently pull
the Connector Cable connector straight from the Generator. Move the CONNECTOR DOOR fully to the right to
cover the THERAPY CONNECTOR and expose the CHARGER CONNECTOR.
Connect the cable coming from the front of the Charger Module to the CHARGER CONNECTOR on the front of
the Generator (see Section 3.1). Ensure that the CHARGER MODULE cable is fully seated into the CHARGER
CONNECTOR. The BATTERY CHARGING indicator will be displayed with a lightning bolt inside the battery
symbol to indicate the battery is now charging (see Step 2 of Section 5.5 for illustration).
Charge the battery for at least twelve hours prior to use. After twelve hours, BATTERY CAPACITY should show
a completely filled in battery symbol (see Step 4 of Section 5.5 for illustration).
NOTE: The battery is self-discharging and requires periodic recharges even during storage so that the battery
will not discharge to an unacceptably low voltage level, which could ultimately result in battery damage.
LBL 60780 rev. B
EN
20 |
P a g e