Step
Step 4 – Confirm Output Test Result
The output test requires approximately 15 seconds to complete.
During this time the THERAPY STATUS indicator will remain on
green. Upon successful completion of this test, the Generator
will sound four beeps. If no error message appears on the display
then this test has passed successfully.
6.1.3

Inspecting the Generator

Physical inspection the Generator on a daily basis is also recommended to help ensure that all components
required for reliable operation are in good condition.
Physical Condition Inspection
Inspect all surfaces of the Generator exterior including the Charger
Module. Confirm there is no visible damage such as cracks or chips in
any component.
Disconnect the Charger Module from the Generator and slide the
CONNECTOR DOOR left and right. Verify it is not damaged, also verify
it is retained in its track and it slides easily left to right. Reconnect the
Charger Module to the Generator.
Inspect the Connector Cable and AC Mains Cable. Confirm there are
not any damaged, split or cracked materials and electrical contacts are
free of extraneous matter.
6.1.4

Cleaning the Generator

Dirt and extraneous matter may be removed from the Generator and Connector Cable using a soft cotton cloth
or a lint free wipe. If needed, use only isopropyl alcohol sparingly as a cleaning agent.
Do not allow any fluids to penetrate the exterior surfaces of the device. Allow equipment to dry thoroughly
before testing or use.
Clean connector areas carefully. Do not attempt to clean interior surfaces of connectors or connector
contacts. In the event a Connector Cable has become contaminated or malfunctions, remove this cable from
use and contact your Shockwave Medical representative.
WARNING
SHOCK OR FIRE HAZARD
Do not immerse any portion of the Generator in water or other fluids. Do not immerse Connector Cables in water or other
fluids. Avoid spilling any fluids on the Generator. Spilled liquids may cause the Generator to perform inaccurately or
malfunction.
Do not clean with solvents or flammable agents as this may cause damage to the Generator and possibly cause harm to the
user
Do not autoclave or sterilize the Generator or Connector Cables as this may cause the Generator to malfunction.

6.2 Monthly Maintenance

There is no specific test or inspection which is recommended to be conducted on a monthly basis in addition to
the tests and inspections included in Section 6.1. However, it is recommended that the shift supervisor or