Technical Data Relating To The Power Supply; Technical Data Related To Electromagnetic Compatibility (Emc) - ARCHEON EOlife Instructions D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 1
IP rating for protection against
IP44
solids, dust and ingress of water
Electromagnetic compatibility
(EMC) according to EN 60601-
The control parameters and threshold values can be obtained from the manufacturer
1-2
BLE module frequency range
2,402 – 2,480 GHz
Signal strength
Max 4 dBm
Certification
Directive 2014/53/UE, FCC, IC, MIC, KC
Shock and vibration resistance
EN 60601-1-12 (category: resistance in an emergency vehicle)
60 x 45 mm
Screen
Resolution 320 x 240 pixels
EN 60601-1:2006/A1:2013/A12:2014
EN 60601-1-2:2015
EN 60601-1-12:2015
EN 62366-1:2015
Applicable standards
EN 62304:2006/A1:2015
ISO 18562-1:2017
ISO 18562-2:2017
ISO 18562-3:2017
Essential performance
Under normal conditions of use, EOlife® must display the bar graph in real time The
requirements
screen must not freeze
± 4% of the actual value measured above 100 mL of tidal volume (Vt)
± 4% of the actual value measured below 100 mL of tidal volume (Vt)
Measurement accuracy
FlowSense® data:
Flow range: ± 250 slm (standard litre per minute)
Dead space: < 10 ml
Update time: 0 5 ms

2. Technical data relating to the power supply

Specifications
Type
Li-ion
Dimensions (W x H x D)
73 mm x 40 mm x 14 mm
Nominal capacity
1,900 mAh (7 03 Wh)
Nominal voltage
3 7 V
Charging voltage
12-24V CV
Charge time (0% to 95%)
Approx 2 hours
Charging temperature
+10°C to 45°C
Storage temperature
At below -20°C and above +25°C the battery run time is reduced
Lifetime
500 charge cycles (> 80% of the initial capacity at 20°C)
IEC 62133
Certifications
UN38 3
UL1642
48
Battery
Specifications
Weight
206 g including accessories
Unit dimensions (L x H x D)
76 x 74 x 44 (± 1 mm)
Cable length
1,800mm (± 10 mm)
Adapters available
AC input voltage
100 V – 240 V (± 10%)
AC input current
Max 250 mA
AC input frequency
60 Hz / 50 Hz
DC output current
3V to 12V 1000 mA
Output protection
Short circuit
Operating conditions
0°C to 25°C
Storage conditions
-25°C to +70°C
Protection class
II
Unit casing materials
ABS plastic, in accordance with standard UL 94V-0

3. Technical data related to electromagnetic compatibility (EMC)

Manufacturer statement and guidance – electromagnetic interference
EOlife® is designed to operate in an electromagnetic environment as defined below The user should ensure that the
device is used in such an environment
Interference measurements
Compliance
Conducted emissions – Mains
terminal disturbance voltage
Class B: 150 kHz to 30 MHz
according to CISPR11
Radiated emissions – Radiation
Class B: 30 MHz to 1 GHz
disturbance according to CISPR11
Harmonic current emissions
Class A 0 kHz – 2 kHz
according to IEC 61000-3-2
Voltage fluctuations/flicker
0 kHz to 2 kHz
according to IEC 61000-3-3
Charger
USA
Europe
U.K.
Australie, NZ
Electromagnetic environment – Guidance
Equipment suitable for use in households and
buildings used for domestic purposes directly
connected to a low voltage power supply
/
/
49