Alarm message linked to the FlowSense® sensor
During ventilation, if the FlowSense® sensor is defective or not properly connected or if a foreign object on the
connector interferes with the signal transmission, an alarm screen is displayed:
FlowSense failure
"FlowSense failure"
You must reconnect a functional FlowSense® sensor as described above to clear the "FlowSense failure" alarm
Note
The "FlowSense failure" alarm screen can only be displayed from the main screen
EOlife® device failure
If there is a device failure, EOlife® will be unusable: the screen switches off (or no longer comes on) and the red LED
in the top right corner of the screen comes on
See "XI Troubleshooting"
d) Standby mode
If there is no user activity (no ventilation or no action on EOlife®) for 3 minutes, the device goes into «standby» mode
and the screen turns off
To turn the screen on again, simply press the physical ON/OFF button and the summary screen will reappear
If EOlife® remains in standby mode for more than 10 minutes, the device will turn itself off completely
2. EOlife® device set-up screen
a) Set-up screen display
1.
Start up EOlife® by pressing the physical ON/OFF button
2. Press normally the physical ON/OFF button to show the summary screen
42
Alarm screen
Summary screen
3. Next, select the [wrench
] to display the EOlife® set-up screen
To set
the date
Choice of
interface
Summary screen
b) Choice of interface
EOlife® offers two interfaces:
o Basic interface: the EOlife® device guides the user by displaying instructions based on how it interprets the
volumes insufflated, tidal volumes and ventilation frequency, without displaying the values of the ventilation
parameters in real time
o Advanced interface: the EOlife® device guides the user by displaying instructions based on how it interprets
the volumes insufflated, tidal volumes and ventilation frequency, but at the same time displays the values of
the ventilation parameters if users wish to interpret them themselves
c) Data transfer
EOlife® records the ventilation parameters in each ventilation cycle and can record for approximately 5 hours This
data can then be transferred via Bluetooth to a compatible device to extract and analyse the data To avoid losing the
data recorded, it is advisable to transfer the data recorded by EOlife® after each use
Note
Transferring data requires the use of a device compatible with EOlife®, not included with the equipment and
accessories sold with EOlife®
Please contact an approved distributor or the manufacturer for more information
Download the data recorded by EOlife® from the set-up screen:
1.
Select [the
icon] The EOlife® Bluetooth
connection is automatically turned on and an
hourglass is displayed until the device detects the
compatible device
2. After a few seconds, a pairing code is displayed on
both devices Check that they are the same:
o Press on [the
icon] ] if the codes are different
and repeat this step
o If they are the same:
3. Press on [the
icon], The devices are paired and
the connection screen is displayed
4. Press on [the
"Data transfer" icon] The transfer
confirmation screen is displayed
To set the time
Language
selection
Access to main
information

Data transfer

Set-up screen
Paired
55%
Update
Data transfer
Remote device
Connection screen
43