2. Using EOlife®
Assemble EOlife®, attaching the FlowSense® sensor and other devices to ventilate the patient (ventilation interface and bag)
See "IX 1 a) Assembly"
EOlife® assembly
1
2
with the FlowSense® sensor
with the ventilation interface
Start up EOlife® by pressing the
Press on the height of the adult
4
5
physical ON/OFF button
patient in cardiopulmonary arrest
Ventilate the patient in
7
cardiopulmonary arrest for as
long as necessary
See "IX 1 b) EOlife®
ventilation method" and
"IX 1 c) Alarm conditions
and alarm signals"
Information area
Medium
30 : 2
Rotate
55%
(advanced interface)
Vi
Vt
Freq.
Vi = insufflated volume
Vt = tidal volume
420 mL
390 mL
11/min
Freq = ventilation frequency
Bar graph
(real-time Vi)
Good Ventilation
Alarm signal
display area
Main screen
(advanced interface)
34
Assembly with the bag
Assembly
3
Choose the type of resuscitation
6
performed
o Small: the patient's height
o 30:2 mode: alternating
is between 140 and 160 cm
between 30 chest
o Medium: the patient's
compressions and 2
height is between 160 and
ventilations
180 cm
o Continuous mode:
o Tall: the patient's height is
continuous ventilation at a
between 180 and 200 cm
frequency of 10 breaths per
minute
Patient's height
Ventilation
55%
Mode
Small
140 / 160 cm
Medium
160 / 180 cm
Tall
180 / 200 cm
Switch off EOlife® by pressing the ON/OFF button, then on
8
the OFF icon or by pressing the ON/OFF button for 3 seconds
Switching off using
the summary screen
IX. Detailed product description and operating instructions
EOlife® is controlled via the touch screen and the physical ON/OFF button
Caution
Avoid using any object that could scratch the screen

1. Using the device

a) Assembly/disassembly
Assembly
55%
Remove the FlowSense® sensor from its packaging, using the notch to tear the antistatic bag to avoid damaging the
sensor
30 : 2
Caution
Check the sensor's expiry date before using The use by date can be found on the label affixed to the bag ( ) Check
Continuous
that the bag is intact before using the FlowSense® sensor If the FlowSense® sensor bag is torn, discard the sensor
and use another one as water or contaminants may have entered the bag
Connect the FlowSense® sensor to the EOlife® unit, taking care not to expose the printed circuit board to dust or
splash it with water
First insert the part of the FlowSense® with a notch into the connector ridge, and then press firmly on the sensor
so that it comes into contact with the connector pins You will hear a click when the connector tab holds the
FlowSense® sensor in place
Check that the FlowSense® sensor is held properly in place before starting up the EOlife® unit
Tab
Disconnect the ventilation bag and interface
9
(mask or endotracheal tube) and unclip
the FlowSense® sensor by pulling the tab
outwards, as shown by the arrow
Notch
Ridge
click
Disconnecting EOlife®