Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 15
EN – ENGLISH

Care and Cleaning

After each use, disassemble all components of the
system and thoroughly clean using a damp cloth
and neutral pH detergent. Remove all traces of
blood and debris and dry completely. The entire
Budde Halo retractor system (REF# A1040) should
be steam sterilized before each use.
NOTE:
Do not disassemble retractor arm assembly
(REF# 438A1011). Loosen tension on the retractor
arm's adjustment knob, which will allow for
ample separation of the individual ball joints for
appropriate cleaning.
Manual Wash
CAUTIONS
• Alkaline and highly acidic detergents and
solutions cause damage to the devices.
• Channels and crevices found on this device
require particular attention during cleaning.
• Pay special attention to the water quality used
throughout reprocessing. Hard water can damage
the surface of the equipment. Avoid using hard
water. Instead use purified water unless otherwise
specified.
Limitations on reprocessing
• Repeated processing has minimal effects on these
devices. Product life is normally determined by
wear and damage due to use.
• It is important to have Integra NeuroSpecialists
perform routine inspections (twice yearly is
recommended). See contact information below.
INSTRUCTIONS
Containment/Transportation
• Follow health care facility protocol for
safe containment and transport to the
decontamination environment.
• It is recommended that devices are cleaned
immediately after use.
Preparation for Cleaning
• Disassembly is required.
Disassembling
Equipment
Disassemble
components as normal
and place back into tray
Cleaning – Manual Equipment: Water, Neutral pH
Detergent, Soft Bristle Brush, Towels
Method
1. Prepare neutral pH enzymatic detergent solution
(e.g. Endozime® AW Triple Plus with APA (Ruhof),
1:128 ratio) according to detergent manufacturer's
instructions using lukewarm tap water.
2. Prepare equipment for soaking by disassembling
removable parts and loosening connections.
3. Rinse equipment in warm water before placing
into bath.
4. Completely soak equipment in water/detergent
solution for 30 minutes maximum.
5. Clean thoroughly with a soft nylon bristle brush.
NOTE – If possible, use a disposable brush.
6. Rinse in warm purified water until all visible
substances and residual detergent are removed.
NOTE – Make sure to give special attention to
hard-to-reach areas.
7. Thoroughly dry equipment with soft clean towels
and use medically compressed air if needed, to
dry channels, crevices and lumens.
8. Inspect the equipment to make sure there is no
visible organic debris or residue from cleaning
agent.
Repeat process if any soil is detected.
Drying
• Products should be dry at this point. If wetness
or excess liquid is detected, dry with a soft clean
towel.
• Medically compressed air can be used if needed.
9
Table 1
Inspecting for
Cleanliness
Before sterilizing look
into tray for any sign of
organic debris.

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Budde haloA1040A1040a

Table des Matières