Disinfection
Further disinfection of device components may be achieved using one of the following methods.
Method 1: Chemical
With device cleaned and still disassembled, wipe all surfaces with sterile gauze moistened with
70% Isopropyl Alcohol (IPA). Be sure to achieve even coverage on all surfaces.
Assure the device stays wet with the 70% IPA for a minimum of 10 minutes.
Dry the device with a sterile lint free cloth or blow dry with filtered pressurized air.
Method 2: Thermal Rinse
A Thermal Disinfection phase maybe added after the rinse cycle as indicated in table 1.

Maintenance and Care:

The Infinity XR2 Base Unit should be returned to its storage case after each use.
NOTE: Ensure all components have completely dried after cleaning before returning to case.
If unit is dropped or mishandled, it should be inspected for damage.
(REF Inspection section of this instruction manual) If damage occurs, do not use;
return the complete device immediately to Integra LifeSciences for inspection.
To ensure proper function and to extend the life and performance of the equipment, Integra
LifeSciences recommends the following:
Recommended Action
Return the device to the Integra LifeSciences Repairs
department for detailed inspection and servicing.
Request that Integra NeuroSpecialists perform routine
inspections of the device.
In the absence of proper care and servicing of the device, negative effects may be seen after repeated
processing over time which may lead to reduced performance.
Contact information: See the Service and Repair section for contact information on how to return your
device for periodic servicing and to request periodic inspections.
See Inspection and /or Service notes section for routine checks to be performed on the device.

Device Disposal

NOTE : Follow hospital procedures for disposal of this device.
CAUTION!
Recommended Frequency
Once / year
Twice / year