Cleaning of the MAYFIELD Infinity XR2 Base Unit
Follow the cleaning instructions outlined in the Cleaning of the MAYFIELD Infinity XR2 Base Unit.
Do Not Steam Sterilize! The carbon Fiber material and plastic components may be
damaged by heat.
Following these steps is recommended:
1. Remove base unit from operating table support rails.
2. Remove Swivel Adaptor, and Link Arm Assembly from base unit
3. The Infinity XR2 Base Unit should be thoroughly cleaned after each use. Scrub each component
with a soft brush and use only a pH-neutral detergent. Clean thoroughly to remove any traces
of blood and/or debris and to prevent such blood or debris from interfering with function or
movement. Rinse thoroughly with clean water.
4. Dry all components with a soft dry towel.
5. After components are totally dry, re-assemble the components. Unit is ready for next use.