Indication For Use

The Infinity XR2 Base Unit is intended to be used to support a patient during diagnostic
examination and/or surgical procedures where a rigid support between surgical table and
headrest, or skull clamp is necessary, positional freedom is required and where X-ray imaging
modalities will be used.
Failure to read and follow instructions furnished in this product insert may result in
skull pin slippage and serious patient injury, such as scalp lacerations, skull fracture,
or even death.
Do not steam sterilize; components may be damaged by high heat leading to device
damage and reduced performance.
The headrest or skull clamp must be securely fastened to the base unit. Failure to
properly position patient and to fully tighten and secure all adjustable portions of
this or any similar device may result in skull pin slippage and serious patient injury,
such as scalp laceration, skull fracture, or even death.
The user must make sure that any threaded connections are secure and
starbursts have meshed (where applicable) after adjustments are complete. Failure
to do so may result in serious patient injury.
Always make sure the base unit is properly secured to the operating table.
The base unit must not be used if the device appears to be damaged or
functioning incorrectly.
Over-tightening the mechanisms adjustment screws may result in damage to the
unit.
Over extending or overloading the base unit may result in unintended
movement, shortened product life and/or damage to the unit.
Do not alter the construction of this device as it may result in serious patient injury.
WARNING!
CAUTION!
5
EN – English