EN – English
Optional Automatic Wash Cycle:
Any deviation from this guideline could result in damage to equipment as well as
improper cleaning results. Parameters given below must be followed exactly in order
to assure proper cleaning and avoid damage to the equipment.
•
Disassemble the device as indicated in section Cleaning of the MAYFIELD Infinity XR2 Base
Unit.
•
Rinse components under warm tap water prior to placing in washer.
•
Arrange device in a way to avoid contact between components.
•
Select instrument cycle and ensure proper programming.
Table 1 Instrument Cycle
Pre-Wash 1
Enzyme Wash
Rinse 1
Thermal Rinse**
* Hot Tap water not to exceed 82.2C
** Optional phase for disinfection of components
•
Remove from washer and dry completely if needed.
•
Inspect all components to ensure there is no visible organic debris or residue from cleaning
agent. Repeat process if any visible debris or residue is detected.
Time in minutes
Phase
04:00
04:00
10:00
Wash 1
00:30
02:00
CAUTION!
Water
Temperature
Cold tap H
Hot tap H
60° C
Hot tap H
82.2° C
Table 1 Instrument Cycle
14
Detergent type &
Concentration
O
N/A
2
Endozime® AW
Triple plus w/
O*
2
A.P.A., 1:128 ratio
Renu-Klenz™,
1:256 ratio
O*
N/A
2
N/A