INSTRUCTIONS FOR USE
• If the unit has been dropped or damaged, before you use the unit, make sure
facility maintenance has examined the unit completely and determined that
it is safe to operate.
• To avoid the risk of electric shock, the unit must only be connected to a supply
mains with protective earth.
• The potential for electrical shock exists with electrical equipment. Failure to
follow facility protocols may cause death or serious injury.
• If the Allen Advance Table is to be used near high-frequency surgical
equipment, cardiac defibrillators, and cardiac defibrillator-monitors, refer to
the product manufacturer's instructions to make sure the equipment will
operate as intended.
• Incorrect use or handling of the power cord may cause damage to the power
cord. If damage has occurred to the power cord or any of its components,
immediately remove the Allen Advance Table from service, and contact the
applicable maintenance persons.
• Do not expose the unit to excessive moisture.
• Fluid spills on to the electronics can cause a hazard. If such a spill occurs,
unplug the unit and remove it from service. When fluid spills occur outside that
seen in normal use, immediately do as follows:
– Unplug the unit from its power source.
– Remove the patient from the unit.
– Clean the fluid spill from the unit.
– Have maintenance examine the unit completely.
– Do not put the unit back into service until it is completely dry, tested, and
determined to be safe to operate.
• Be aware of these pinch points on the Allen Advance Table:
– Between the caster and base skins
– Between an H-bracket and the column
– Between the head of the cross pin and the end cap during an extreme tilt
and Trendelenburg, or Reverse Trendelenburg position
– Between the cross pin drop lock and the bowtie
– Between an H-bracket and the column during the removal of a cross pin
• Do not service the unit while a patient is on the unit. Consult with facility
caregivers to make sure that a patient is not on the unit before you service the
unit. Patient injury can occur.
Document Number: 80028190
Version: E
Page 62
English
Issue Date: 16 MAY 2022
Ref Blank Template: 80025117 Ver. F